Gas Filter
Gas Chromatograph
Enables tool-free column connection. Septum replacement is also possible.
Sealing mechanism | ClickTek nut |
---|---|
Injection mode | Split, Splitless, High Pressure Injection |
Split ratio | Up to 9999.9:1 |
Pressure range | 0 to 1035 kPat |
---|---|
Maximum operating temperature | 450 ℃ |
Supports capillary columns from 50 μm to 530 μm I.D. / Electronic septum purge comes built-in / Gas saver mode reduces gas consumption / Deactivation process (Option)
The FlD shows good response for all organic compounds.
It is characterized by high robustness and a wide linear dynamic range
Minimum Detectable Quantity (MDQ) |
< 1.5 pg C/s (dodecane) |
---|---|
Linear dynamic range | 1 × 107 (± 10 %) |
Max acquisition rate | 2 ms (500 Hz) |
Maximum operating temperature |
450 ℃ |
Automatic extinguishing detection/reignition |
Standard-compatible |
Flow rate settings | Makeup (N2 or He) : 0 to 100 mL/min H2 : 0 to 100 mL/min Air : 0 to 1000 mL/min |
---|
The detector of choice for trace-level analysis of phosphorus (P), sulfur (S) and tin (Sn) compounds.
Minimum Detectable Quantity (MDQ) |
< 55.0 fg P/s (tributyl phosphate) |
---|---|
Minimum Detectable Quantity (MDQ) |
< 2.5 pg S/s (dodecanethiol) |
Easy switching of luminous filters |
P, S, Sn |
Dynamic range | Tributyl phosphate (P) : 1 × 104 Dodecanethiol (S) : 1 × 103 |
Max acquisition rate | 2 ms (500 Hz) |
Max operating temperature | 450 ℃ |
Flow rate settings | H2 : 0 to 250 mL/min Air : 0 to 1000 mL/min |
---|
The detector of choice for trace-level analysis of electrophilic compounds such as halogenated, organometallic and nitro compounds.
Minimum Detectable Quantity (MDQ) |
< 4.0 fg/s (Lindane) |
---|---|
Dynamic range | 1 × 105 |
Max acquisition rate | 2 ms (500 Hz) |
Max operating temperature | 400 ℃ |
---|---|
Flow rate settings | ECD gas (N2, Ar) : 0 to 200 mL/min |