Print this page

 
Shimadzu - Solutions for Science since 1875 Shimadzu France
Global Links
Vous êtes ici: >> Skip Navigation LinksHome > Products > Spectroscopy > AAS > AA-7000 > Specifications

Spectroscopy

AA - 7000
Specifications

Optics
Wavelength range 185.0 to 900.0 nm
Monochromator Aberration-corrected Czerny-Turner mounting
Bandwidth 0.2, 0.7, 1.3, 2.0L nm (4-step automatic switching)
Detector Photomultiplier tube
Optics
  • Flame: optical double beam
  • Furnace: high-throughput single beam
Background correction
  • BGC-SR (high-speed self-reversal method)
  • BGC-D2 (D2 lamp method)
Number of HC lamps 6-lamp turret, 2 lamps simultaneously lit (1 for measurement, 1 warming up for next measurement)
Lamp mode EMISSION, NON-BGC, BGC-D2, BGC-SR
Data processing
Software requirements Microsoft Windows Vista Business / XP Professional
Parameter setting Wizard method
Measurement mode Flame continuous method, flame single drop method, furnace method
Concentration computation mode
  • Calibration curve method (select primary, secondary, tertiary)
  • Standard addition method, simple standard addition method (primary expression)
Repeat analysis Up to 20 repetitions. Mean value, standard deviation (SD) and coefficient of variation (RSD) displayAutomatic exclusion of deviant values by setting SD and %RSD
Baseline correction Automatic correction of baseline drift by offset correction in peak height / peak area modes.
Signal processing segment setting Signal processing segments can be changed in peak height / peak area modes.
Sensitivity correction Automatic calibration curve correction function using sensitivity monitoring
Analog output 2 channels (atomic absorption/energy signal, background signal)Output range: 5.0, 2.5, 1.25, 0.625 Abs./V (each settable in 4 stages) Fixed at 1 V F.S. in EMISSION mode.
Tabular data processing Final concentration calculations based on sampled volume, dilution rate, fixed volume, and factor inputs
Recall of parameters Template functions available
Procedure/result display MRT (Measurement Results Table) worksheet
Report generation Summary report available
QA/QC Select whether to continue or discontinue measurements based on results of evaluation on coefficient of correlation, %RSD, ICVŸICB, CCVŸCCB, PB, LCS, SPK, PDS, and DUP.
Re-analysis
  • Select whether on not to conduct re-analysis.
  • Automatic dilution and re-analysis of unknown samples via autosampler (flame single drop method, furnace method)
Digital recording
  • Management by login ID and password
  • Control user access authority by user level
  • Log recordŸAudit trail
  • Electronic signatures
Power requirements 120 VAC (English Ver.), 220 VAC (Chinese Ver.), or 230 VAC (English Ver.), 230 VA, 50/60 HzPower is required separately for the personal computer.
Dimensions and weight AA-7000F: 700 W x 588 D x 714 H mm, 72 kgAA-7000G: 700 W x 580 D x 538 H mm, 65 kg(Protruding parts and optional equipment are not included.)
Ambient temperature / humidity 10 to 35 °C, 20 to 80% (less than 70% when temperature is higher than 30 °C)


Shimadzu France | 65 Avenue du Général de Gaulle | 77420 Champs sur Marne
Pour plus d’information contactez nous par téléphone au 01 60 95 10 10, par fax au 01 60 06 51 66, par mail sur shimadzu@shimadzu.fr
ou posez vos questions dans notre rubrique contact


© Shimadzu Europa GmbH | Albert-Hahn-Str. 6-10 | 47269 Duisburg | Tel: 0203-76870 | Fax: 0203-766625