With all these features the system is ready for operation
whichever sample should be analyzed, aqueous solutions or organic solvents in the
concentration range of % down to µg/l. The optimization of all system parameters
takes care on maximum energy and highest sensitivity, that means with an automatic
flame optimization or burner positioning function the instrument meets every sample.
The quantitation of multielements in ultra trace concentration levels is done on
the AA-6300 in combination with a digital controlled Graphitefurnace GFA-EX7. In
this combination the instruments works on a maximum energy level and electronics
double beam stability. The digital furnace with high sensitivity and intelligent
temperature control results in detection limits (acc. to DIN 51401/1)which are almost
out of imagination. The detection limit for lead in furnace measured according to
3 sigma is 0,085 µg/L using an injection volume of 20 µL, resulting
in an absolute value of 1,7 pg. Furthermore this furnace concept guarantees extremely
long graphite tube lifetimes (2000 shots at 2800 °C and an HNO3 acid
concentration of 0,1 % in aqueous solution). Table 1 shows typical detection limits
of elements in graphite furnace operation using an injection volume of 20 µL.